User manual
Remote Control
R&S
®
RTO
748User Manual 1316.0827.02 ─ 06
TRIGger<m>:PARallel:DATatoclock:CSOurce:EDGE <ClockEdge>
Sets the edge of the clock signal. The crossing of the clock edge and the logical threshold
defines the time reference point for the setup and hold time measurement.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger<m>:PARallel:DATatoclock:STIMe <SetupTime>
Sets the minimum time before the clock edge while data should be stable and not change
its state.
See also: "Setup time" on page 337
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<SetupTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
TRIGger<m>:PARallel:DATatoclock:HTIMe <HoldTime>
Sets the minimum time after the clock edge while data should be stable and not change
its state.
See also: "Hold time" on page 337
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<HoldTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s