User manual

Remote Control
R&S
®
RTO
747User Manual 1316.0827.02 ─ 06
Timeout Trigger
See also:
TRIGger<m>:PARallel:TIMeout:EXPRession[:DEFine] on page 744
TRIGger<m>:PARallel:TIMeout:RANGe...........................................................................747
TRIGger<m>:PARallel:TIMeout:TIME...............................................................................747
TRIGger<m>:PARallel:TIMeout:RANGe <TimeoutMode>
Sets the state condition.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<TimeoutMode> HIGH | LOW | EITHer
HIGH
The level of a digital channel stays above the threshold, or the
logical expression for LOGic trigger source is true.
LOW
The level of a digital channel stays below the threshold, or the
logical expression for LOGic trigger source is false.
EITHer
The signal state remains unchanged.
*RST: HIGH
TRIGger<m>:PARallel:TIMeout:TIME <Time>
Defines the time limit for the timeout at which the instrument triggers.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<Time> Range: 100E-12 to 10000
Increment: 100E-9
*RST: 100E-9
Default unit: s
Data2Clock Trigger
See also:
TRIGger<m>:PARallel:DATatoclock:CSOurce[:VALue] on page 744
TRIGger<m>:PARallel:DATatoclock:CSOurce:EDGE........................................................748
TRIGger<m>:PARallel:DATatoclock:STIMe......................................................................748
TRIGger<m>:PARallel:DATatoclock:HTIMe......................................................................748
Command Reference