746User Manual 1316.0827.02 ─ 06
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin | OUTSide
Pulses inside or outside a given time range. The time range Width
±
Delta is defined by TRIGger<m>:PARallel:WIDTh:WIDTh
and TRIGger<m>:PARallel:WIDTh:DELTa.
SHORter | LONGer
Pulses shorter or longer than a given width defined by
TRIGger<m>:PARallel:WIDTh:WIDTh
*RST: WITHin
TRIGger<m>:PARallel:WIDTh:WIDTh <Width>
Sets the limit for the pulse width.
The effect depends on TRIGger<m>:PARallel:WIDTh:RANGe.
●
For the ranges SHORter and LONGer, the width defines the maximum and minimum
pulse width, respectively.
●
For the ranges WITHin and OUTSide, the width defines the center of a range which
is defined by the limits TRIGger<m>:PARallel:WIDTh:DELTa.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<Width> Range: 200E-12 to 10000
Increment: 200E-9
*RST: 5E-9
Default unit: s
TRIGger<m>:PARallel:WIDTh:DELTa <WidthDelta>
Defines a range around the given width value. the setting is relevant if TRIGger<m>:
PARallel:WIDTh:RANGe is set to WITHin or OUTSide. The width is set with
TRIGger<m>:PARallel:WIDTh:WIDTh.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 600E-12
*RST: 0
Default unit: s