User manual
Remote Control
R&S
®
RTO
739User Manual 1316.0827.02 ─ 06
DIGital<m>:DESKew <Deskew>
Sets an individual delay for each digital channel to time-align it with other digital channels.
The deskew value compensates delays that are known from the circuit specifics or
caused by the different length of cables.
Suffix:
<m>
.
0..15
Number of the digital channel
Parameters:
<Deskew> Default unit: s
DIGital<m>:THReshold
<Value>
Sets the logical threshold for the indicated digital channel.
Suffix:
<m>
.
0..15
Number of the digital channel
Parameters:
<Value> Voltage value
Firmware/Software: V 1.30
DIGital<m>:HYSTeresis <Hysteresis>
Defines the size of the hysteresis to avoid the change of signal states due to noise.
Suffix:
<m>
.
0..15
Number of the digital channel
Parameters:
<Hysteresis> MAXIMUM | ROBUST | NORMAL
MAXIMUM
Maximum value that is possible and useful for the signal and its
settings
ROBUST
Different hysteresis values for falling and rising edges to avoid an
undefined state of the trigger system.
NORMAL
The instrument sets a value suitable for the signal and its settings.
Firmware/Software: V 1.30
16.2.15.2 Parallel Bus Configuration
The following commands configure the four parallel buses of R&S RTO-B1.
Command Reference