User manual

Remote Control
R&S
®
RTO
733User Manual 1316.0827.02 ─ 06
Return values:
<FrameState> OK | FSS | BSS | FES | INDicator | HCRCerror | CRCerr | LENER |
WAKeup | MTS | INSufficient
OK: the frame is valid.
FSS: Frame Start Sequence after TSS is missing.
BSS: Byte Start Sequence is missing.
FES: error in the Frame End Sequence.
INDicator: Error in indicator bits.
HCRCerror: Header CRC is not valid.
CRCerr: Payload CRC is not valid.
LENER: Unexpected length of the frame.
WAKeup: frame contains a wakeup pattern.
MTS: frame contains a MTS or CAS symbol
INSufficient: The frame is not completely contained in the acqui-
sition. The acquired part of the frame is valid.
*RST:
OK
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:STARt?
BUS<m>:FLXRay:FRAMe<n>:STOP?
Return the start time and stop time of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<FrameStart>
<FrameStop>
Time
Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:FLXRay:FRAMe<n>:TYPE?
Returns the frame type of the selected frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Command Reference