User manual

Remote Control
R&S
®
RTO
696User Manual 1316.0827.02 ─ 06
Return values:
<FrameType> DATA | REMote | ERR | OVLD
Data, remote, error or overload frame.
*RST: DATA
Usage: Query only
BUS<m>:CAN:FRAMe<n>:DATA?
Returns the data bytes of the specified frame.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<Data>
Comma-separated list of values. To set the value format, use
FORMat:BPATtern.
Usage: Query only
BUS<m>:CAN:FRAMe<n>:ACKState?
BUS<m>:CAN:FRAMe<n>:CSSTate?
BUS<m>:CAN:FRAMe<n>:DLCState?
BUS<m>:CAN:FRAMe<n>:IDSTate?
Return the states of following parts of a message
ACKState: state of acknowledgement field
CSSTate: state of checksum field (CRC)
DLCState: state of data length code
IDSTate: identifier state
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the frame.
Return values:
<State> OK | UNDF
UNDF: Undefined
*RST: OK
Usage: Query only
Command Reference