693User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<AckError> ON | OFF
*RST: ON
TRIGger<m>:CAN:BITSterror <BitStuffingError>
Triggers if astuff error occurs - when the 6th consecutive equal bit level in the mentioned
fields is detected.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<BitStuffingError> ON | OFF
*RST: ON
TRIGger<m>:CAN:CRCerror <ChecksumError>
Triggers on CRC errors. A CRC error occurs when the CRC calculated by the receiver
differs from the received value in the CRC sequence.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<ChecksumError> ON | OFF
*RST: ON
TRIGger<m>:CAN:FORMerror <FormError>
Triggers when a fixed-form bit field contains one or more illegal bits.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<FormError> ON | OFF
*RST: ON
Command Reference