Parameters:
<Endianness> BENDian | LENDian
BENDian
Big endian, most significant byte first.
LENDian
Little endian, least significant byte first.
*RST: BENDian
TRIGger<m>:CAN:DLCCondition <DLCOperator>
Operator to set the data length code for triggering on CAN data. For big endian transfer
direction, you can trigger on a number of bytes less than the DLC of the frame, that means,
on the beginning of the data pattern. For little endian transfer direction, the exact number
of data bytes in the frame must be set. The number of data bytes to be found is set with
TRIGger<m>:CAN:DLC on page 692.
Example: DLC ≥ 2. The frame has at least two bytes, and you trigger on the data of the
first two bytes.
See also: TRIGger<m>:CAN:BORDer on page 691.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DLCOperator> EQUal | GETHan
For little endian transfer direction, EQUal must be set.
*RST: GETHan
TRIGger<m>:CAN:DLC <WordCount>
Sets the Data Length Code, the number of data bytes to be found. For complete definition,
set also the operator with TRIGger<m>:CAN:DLCCondition on page 692.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<WordCount> Range: 1 to 8
Increment: 1
*RST: 1
TRIGger<m>:CAN:ACKerror <AckError>
Triggers when the transmitter does not receive an acknowledgment - a dominant bit dur-
ing the Ack Slot.
The trigger type has to be set before: TRIGger<m>:CAN:TYPE to ERRC.