For data and remote frames, the identifier format has to be set with TRIGger<m>:
CAN:ITYPe.
See also: "Frame type" on page 290
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<FrameType> ANY | DATA | REMote | ERRor | OVERload
*RST: ANY
TRIGger<m>:
CAN:ITYPe <IdentifierType>
Selects the format of data and remote frames.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<IdentifierType> ANY | B11 | B29
B11
11 bit identifier (standard format). The instrument triggers on the
sample point of the IDE bit.
B29
29 bit identifier (extended format). The instrument triggers on the
sample point of the RTR bit.
*RST: ANY
TRIGger<m>:CAN:ICONdition <IdentifierOperator>
Sets the operator to set a specific identifier or an identifier range.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only