Parameters:
<Technology> V25 | V3 | V2 | V0 | MAN
V25
2.5 Volt (CMOS 5.0 V)
V3
3.0 Volt (CAN_H HS / CAN_L LS)
V2
2.0 Volt (CAN_L HS / CAN_H LS)
V0
Ground
MAN
Manual setting of user-defined values with BUS<m>:CAN:DATA:
THReshold.
*RST:
V25
BUS<m>:
CAN:BITRate <Bitrate>
Sets the number of transmitted bits per second.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Bitrate> Range: 100 to 5000000
Increment: 1
*RST: 100000
Default unit: bps
BUS<m>:CAN:SAMPlepoint <SamplePoint>
Sets the position of the sample point within the bit in percent ot the nominal bit time.
Alternatively, you can set the sample point with BUS<m>:CAN:T1Segment and
BUS<m>:CAN:T2Segment .
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SamplePoint> Range: 10 to 90
Increment: 5
*RST: 50
Default unit: %