User manual
Remote Control
R&S
®
RTO
681User Manual 1316.0827.02 ─ 06
TRIGger<m>:UART:DPOSition <DataPosition>
Sets the number of words before the first word of interest. These offset words are ignored.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataPosition> Number of words
Range: 0 to 32767
Increment: 1
*RST: 0
Usage: Asynchronous command
TRIGger<m>:UART:DPTO <DataPositionTo>
Defines the last word of interest, if TRIGger<m>:UART:DPOPerator defines a position
range.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataPositionTo> Range: 0 to 32767
Increment: 1
*RST: 0
Usage: Asynchronous command
TRIGger<m>:UART:FCONdition <DataOperator>
Selects the operator for the data pattern (TRIGger<m>:UART:DATA).
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataOperator> EQUal | NEQual
*RST: EQUal
TRIGger<m>:UART:DATA <Data>
Specifies the data pattern to be found on the specified trigger source, in binary or hex
format. Enter the words in msb first bit order.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Command Reference