BUS<m>:SPI:SCLK:THReshold <SCLKThreshold>
BUS<m>:SPI:MISO:THReshold <MISOThreshold>
BUS<m>:SPI:MOSI:THReshold <MOSIThreshold>
BUS<m>:SPI:SSELect:THReshold <SSThreshold>
Set user-defined threshold values for the clock, MISO, MOSI and slave select lines.
Alternatively, you can set the thresholds according to the signal technology with
BUS<m>:SPI:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SSThreshold> User-defined value
Range: -12 to 12
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:SPI:FRCondition <FrameCondition>
Defines the start of a frame. A frame contains a number of successive words, at least
one word.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<FrameCondition> SS | CLKTimeout
SS
Start and end of the frame is defined by the active state of the slave
select signal, see BUS<m>:SPI:SSELect:POLarity.
CLKTimeout
Defines a timeout on the clock line SCLK as limiter between two
frames. The timeout condition is used for SPI connections without
an SS line.
*RST: SS
BUS<m>:SPI:TIMeout <ClockTimeout>
Defines a timeout on the clock line SCLK as limiter between two frames. The timeout
condition is used for SPI connections without an SS line.
Suffix:
<m>
.
1..4
Selects the serial bus.