User manual

Acquisition and Setup
R&S
®
RTO
52User Manual 1316.0827.02 ─ 06
Make sure that the correct channel tab is selected.
Show channel
Switches the channel signal on or off. The signal icon appears on the signal bar. The
waveform of the last acquisition is displayed in the diagram.
SCPI command:
CHANnel<m>:STATe on page 436
Use skew offset
If enabled, the "Skew offset" value is used for compensation. This improves horizontal
and trigger accuracy.
SCPI command:
CHANnel<m>:SKEW:MANual on page 452
Skew offset
Sets an delay value, that is known from the circuit specifics but cannot be compensated
by the instrument automatically. It affects only the selected input channel.
SCPI command:
CHANnel<m>:SKEW:TIME on page 452
2.3.5.3 AUX OUT
1 GHz Reference ON
Enables the 1 GHz reference signal and sends it to the AUX OUT connector at the front
panel. The signal is required for performance test to measure the frequency internal cal-
ibration signal.
Reference for Acquisition and Setup