TRIGger<m>:I2C:DPOSition <DataPosition>
Sets the number of data bytes before the first byte of interest. These bytes are ignored.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataPosition> The index 0 is associated with the first data byte.
Range: 0 to 4095
Increment: 1
*RST: 0
TRIGger<m>:
I2C:DPTO <DataPositionTo>
Defines the last byte of interest, if TRIGger<m>:I2C:DPOPerator is set to RANGe.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataPositionTo> Range: 0 to 4095
Increment: 1
*RST: 0
TRIGger<m>:I2C:DCONdition <DataOperator>
Sets the operator to set a specific data value or a data range.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange |
OORange
*RST: EQUal
TRIGger<m>:I2C:DMIN <Data>
Specifies the data bit pattern, or sets the the start value of a data pattern range.. Enter
the bytes in msb first bit order. The maximum pattern length is 64 bit. Waveform data is
compared with the pattern byte-by-byte.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Command Reference