User manual

Remote Control
R&S
®
RTO
649User Manual 1316.0827.02 ─ 06
Parameters:
<AddressNack> ON | OFF
*RST: ON
Usage: Asynchronous command
TRIGger<m>:I2C:DWNack <DataWriteNack>
Triggers if a date acknowledge bit is missing - the addressed slave does not accept the
data.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataWriteNack> ON | OFF
*RST: ON
Usage: Asynchronous command
TRIGger<m>:I2C:DRNack <DataReadNack>
Triggers on the end of the read process when the master reads data from the slave. This
Nack is sent according to the protocol definition, it is not an error.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only
Parameters:
<DataReadNack> ON | OFF
*RST: ON
Usage: Asynchronous command
TRIGger<m>:I2C:AMODe <AddressType>
Sets the address length. The setting affects the address input with TRIGger<m>:I2C:
ADDRess and TRIGger<m>:I2C:ADDTo.
Suffix:
<m>
.
1..3
Event in a trigger sequence: 1 = A-event only