User manual
Remote Control
R&S
®
RTO
646User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SCLThreshold> User-defined clock threshold
Range: -12 to 12
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:I2C:SDA:THReshold <SDAThreshold>
Sets a user-defined threshold value for the data line.
Alternatively, you can set the threshold according to the signal technology with
BUS<m>:I2C:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SDAThreshold> User-defined data threshold
Range: -12 to 12
Increment: 0.1
*RST: 0
Default unit: V
BUS<m>:I2C:TECHnology <Technology>
Sets the threshold voltage clock and data lines as defined for various signal technologies.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Technology> V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MAN
V15 | V25 | V165 | V125 | V09 | V38 | V20 | V0
1.5 V, 2.5 V, 1.65 V ... respectively
VM13
-1.3 V (negative value)
MAN
Manual setting of user-defined values with BUS<m>:I2C:SCL:
THReshold and BUS<m>:I2C:SDA:THReshold.
*RST: V165
Usage: SCPI confirmed
Command Reference