User manual

Remote Control
R&S
®
RTO
634User Manual 1316.0827.02 ─ 06
Parameters:
<Stop> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0.01
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:GATE:RELative:STARt <SearchName>,<RelativeStart>
SEARch:GATE:RELative:STARt? <SearchName>
Defines the starting value for the gate.
Parameters:
<RelativeStart> Range: -100E+24 to 100E+24
Increment: 0.1
*RST: 0
Default unit: %
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:GATE:RELative:STOP <SearchName>,<RelativeStop>
SEARch:GATE:RELative:STOP? <SearchName>
Defines the end value for the gate.
Parameters:
<RelativeStop> Range: -100E+24 to 100E+24
Increment: 0.1
*RST: 100
Default unit: %
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:GATE:ZCOupling <SearchName>,<ZoomCoupling>
SEARch:GATE:ZCOupling? <SearchName>
If enabled, the gate area is defined identically to the zoom area for the zoom diagram.
Parameters:
<ZoomCoupling> ON | OFF
*RST: OFF
Parameters for setting and query:
<SearchName> Search definition
Command Reference