SEARch:TRIGger:WINDow:TIMerange <SearchName>,<TimeRangeMode>
SEARch:TRIGger:WINDow:TIMerange? <SearchName>
Selects how the time limit of the window is defined. Time conditioning is available for the
vertical conditions "WITHin" and "OUTSide" (see SEARch:TRIGger:WINDow:RANGe
on page 629).
Parameters:
<TimeRangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers if the signal stays inside or outside the vertical window
limits at least for the time Width - Delta and for Width + Delta at
the most.
OUTSide
"Outside" is the opposite definition of "Within". The instrument
triggers if the signal stays inside or outside the vertical window
limits for a time shorter than Width - Delta or longer than Width +
Delta.
SHORter
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
*RST:
WITHin
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:WINDow:WIDTh <SearchName>,<Width>
SEARch:TRIGger:WINDow:WIDTh? <SearchName>
For the ranges "Within" and "Outside", the width defines the center of a time range which
is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time lapse,
respectively.
The range is defined using SEARch:TRIGger:WINDow:RANGe.
Parameters:
<Width> Range: 100E-12 to 864
Increment: 100E-9
*RST: 5E-9
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed