SEARch:TRIGger:LEVel:WINDow:UPPer.........................................................................631
SEARch:TRIGger:WINDow:DELTa <SearchName>,<WidthDelta>
SEARch:TRIGger:WINDow:DELTa? <SearchName>
Defines a range around the "Width" value (see SEARch:TRIGger:WINDow:WIDTh
on page 630).
Parameters:
<WidthDelta> Range: 0 to 432
Increment: 500E-12
*RST: 0
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:WINDow:RANGe <SearchName>,<RangeMode>
SEARch:TRIGger:WINDow:RANGe? <SearchName>
Selects how the signal run is compared with the window.
Parameters:
<RangeMode> ENTer | EXIT | WITHin | OUTSide
ENTer
Triggers when the signal crosses the upper or lower level and thus
enters the window made up of these two levels.
EXIT
Triggers when the signal leaves the window.
WITHin
Triggers if the signal stays between the upper and lower level for
a specified time. The time is defined in various ways by the
SEARch:TRIGger:WINDow:TIMerange command.
OUTSide
Triggers if the signal stays above the upper level or below the
lower level for a specified time. The time is also defined by the
SEARch:TRIGger:WINDow:TIMerange command.
*RST:
ENTer
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
Command Reference