User manual

Remote Control
R&S
®
RTO
626User Manual 1316.0827.02 ─ 06
Parameters for setting and query:
<Key> Search definition
<SignalSource> NONE | C1W1 | C1W2 | C1W3 | C2W1 | C2W2 | C2W3 | C3W1 |
C3W2 | C3W3 | C4W1 | C4W2 | C4W3 | M1 | M2 | M3 | M4 | R1 |
R2 | R3 | R4
Source of the search, see chapter 16.2.2.2, "Waveform Parame-
ter", on page 419
*RST: NONE
Usage: SCPI confirmed
Timeout
SEARch:TRIGger:TIMeout:RANGe..................................................................................626
SEARch:TRIGger:TIMeout:TIME.....................................................................................626
SEARch:TRIGger:TIMeout:RANGe <SearchName>,<TimeoutMode>
SEARch:TRIGger:TIMeout:RANGe? <SearchName>
Selects the relation of the signal level to the trigger level:
Parameters:
<TimeoutMode> HIGH | LOW | EITHer
HIGH
The signal level stays above the trigger level.
LOW
The signal level stays below the trigger level.
EITHer
The signal level stays above or below the trigger level.
*RST: HIGH
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:TIMeout:TIME <SearchName>,<Time>
SEARch:TRIGger:TIMeout:TIME? <SearchName>
Defines the time limit for the timeout at which the instrument triggers.
Parameters:
<Time> Range: 100E-12 to 864
Increment: 100E-9
*RST: 100E-9
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Command Reference