User manual

Remote Control
R&S
®
RTO
624User Manual 1316.0827.02 ─ 06
Parameters:
<TimeDelta> Range: 0 to 10
Increment: 100E-9
*RST: 0
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:SLEWrate:RANGe <SearchName>,<RangeMode>
SEARch:TRIGger:SLEWrate:RANGe? <SearchName>
Selects how the time limit for the slew rate is defined. The time measurement starts when
the signal crosses the first trigger level - the upper or lower level depending on the
selected slope - and stops when the signal crosses the second level.
Parameters:
<RangeMode> INSRange | OUTRange | LTHan | GTHan
INSRange
Triggers on slew rates inside a given time range. The range is
defined by "Slew rate" and "±Delta".
OUTRange
Triggers on slew rates outside a given time range. The range def-
inition is the same as for "Within" range.
LTHan
Triggers on slew rates shorter than the given "Slew rate" limit.
GTHan
Triggers on slew rates longer than the given "Slew rate" limit.
*RST:
GTHan
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:SLEWrate:SLOPe <SearchName>,<Slope>
SEARch:TRIGger:SLEWrate:SLOPe? <SearchName>
Selects the edge type.
Parameters:
<Slope> POSitive | NEGative | EITHer
See chapter 16.2.2.3, "Slope Parameter", on page 420.
*RST: POSitive
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed