Parameters:
<Mode> ANY | LONGer | SHORter | WITHin | OUTSide
ANY
Triggers on all runts fulfilling the level condition, without time lim-
itation.
LONGer
Triggers on runts longer than the given "Runt width".
SHORter
Triggers on runts shorter than the given "Runt width".
WITHin
Triggers if the runt length is inside a given time range. The range
is defined by "Runt width" and "±Delta".
OUTSide
Triggers if the runt length is outside a given time range. The range
definition is the same as for "Within" range.
*RST:
ANY
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:RUNT:WIDTh <SearchName>,<Width>
SEARch:TRIGger:RUNT:WIDTh? <SearchName>
For the ranges "Shorter" and "Longer", the runt width defines the maximum and minimum
pulse width, respectively.
For the ranges "Within" and "Outside", the runt width defines the center of a range which
is defined by "±Delta".
The range is defined using SEARch:TRIGger:RUNT:RANGe on page 621.
Parameters:
<Width> Range: 100E-12 to 864
Increment: 100E-9
*RST: 5E-9
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:LEVel:RUNT:LOWer <Key>, <SignalSource>, <Value>
SEARch:TRIGger:LEVel:RUNT:LOWer? <Key>, <SignalSource>
Sets the lower voltage threshold.
Parameters:
<Value> Voltage value
Command Reference