SEARch:TRIGger:DATatoclock:CEDGe <SearchName>,<ClockEdge>
SEARch:TRIGger:DATatoclock:CEDGe? <SearchName>
Sets the edge of the clock signal to define the time reference point for the setup and hold
time.
Parameters:
<ClockEdge> POSitive | NEGative | EITHer
See chapter 16.2.2.3, "Slope Parameter", on page 420.
*RST: POSitive
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:
TRIGger:DATatoclock:CLEVel <SearchName>,<ClockLevel>
SEARch:TRIGger:DATatoclock:CLEVel? <SearchName>
Sets the voltage level for the clock signal. Both this command and SEARch:TRIGger:
DATatoclock:CEDGe define the starting point for calculation of the setup and hold time.
Parameters:
<ClockLevel> Range: -10 to 10
Increment: 1E-3
*RST: 0
Default unit: V
Parameters for setting and query:
<SearchName> Search definition
Usage: SCPI confirmed
SEARch:TRIGger:DATatoclock:CSOurce <SearchName>,<ClockSource>
SEARch:TRIGger:DATatoclock:CSOurce? <SearchName>
Selects the waveform used for the clock signal.
Parameters:
<ClockSource> NONE | C1W1 | C1W2 | C1W3 | C2W1 | C2W2 | C2W3 | C3W1 |
C3W2 | C3W3 | C4W1 | C4W2 | C4W3 | M1 | M2 | M3 | M4 | R1 |
R2 | R3 | R4
Source of the clock signal, see chapter 16.2.2.2, "Waveform
Parameter", on page 419
*RST: C1W1
Parameters for setting and query:
<SearchName> Search definition name
Usage: SCPI confirmed
Command Reference