User manual
Remote Control
R&S
®
RTO
607User Manual 1316.0827.02 ─ 06
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
MTESt:WFMRescale:YPOSition <MaskTestName>,<VertPosi>
MTESt:WFMRescale:YPOSition? <MaskTestName>
Moves the mask vertically within the display.
Parameters:
<VertPosi> Range: -100E+24 to 100E+24
Increment: 0.05
*RST: 0
Default unit: div
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
MTESt:
WFMRescale:YSTRetch <MaskTestName>,<VerticalStretch>
MTESt:WFMRescale:YSTRetch? <MaskTestName>
Sets the vertical scaling to stretch the mask in y-direction. The scaling axis is the hori-
zontal line through the lowest value of the lower mask limit.
Parameters:
<VerticalStretch> Scale factor in %
Range: 10 to 1000
Increment: 1
*RST: 100
Default unit: %
Parameters for setting and query:
<MaskTestName> String containing the name of the mask test
16.2.12.4 Event Actions
MTESt:ONViolation:BEEP...............................................................................................607
MTESt:ONViolation:STOP..............................................................................................608
MTESt:ONViolation:PRINt...............................................................................................608
MTESt:ONViolation:SAVewaveform.................................................................................608
MTESt:ONViolation:BEEP <MaskTestName>,<Beep>
MTESt:ONViolation:BEEP? <MaskTestName>
Generates a beep sound for the specified event.
Parameters:
<Beep> NOACtion | SUCCess | VIOLation
See chapter 16.2.2.5, "Event Parameter", on page 421
*RST: NOACtion
Command Reference