605User Manual 1316.0827.02 ─ 06
MTESt:SEGMent:RESCale:XOFFset <MaskTestName>, <MaskSegmIdx>,<OffsetX>
MTESt:SEGMent:RESCale:XOFFset? <MaskTestName>, <MaskSegmIdx>
Moves the mask segment horizontally. The specified offset is added to the x-values of all
points of the selected mask segment.
Only takes effect after the MTESt:SEGMent:RESCale:RECalculate command.
Parameters:
<OffsetX> Range: -50 to 50
Increment: 1E-9
*RST: 0
Default unit: s
Parameters for setting and query:
<MaskTestName>
<MaskSegmIdx> Index of the mask segment in the mask definition
Firmware/Software: V 1.25
MTESt:SEGMent:RESCale:YOFFset <MaskTestName>, <MaskSegmIdx>,<OffsetY>
MTESt:SEGMent:RESCale:YOFFset? <MaskTestName>, <MaskSegmIdx>
Moves the mask segment vertically. The specified offset is added to the y-values of all
points of the selected mask segment.
Only takes effect after the MTESt:SEGMent:RESCale:RECalculate command.
Parameters:
<OffsetY> Range: -1000 to 1000
Increment: 1E-6
*RST: 0
Default unit: V
Parameters for setting and query:
<MaskTestName>
<MaskSegmIdx> Index of the mask segment in the mask definition
16.2.12.3 Mask Definition: Waveform Mask
The chapter contains commands required for the definition of waveform masks -
MTESt:CTYPe is set to WFML.
MTESt:REFWfm <MaskTestName>,<Source>
MTESt:REFWfm? <MaskTestName>
Sets the reference waveform from which the mask is created.
The reference waveform can be created before, or loaded from a file with REFCurve
commands, see chapter 16.2.11.1, "Reference", on page 590.