User manual

Remote Control
R&S
®
RTO
596User Manual 1316.0827.02 ─ 06
Parameters:
<Offset> Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0
Default unit: s
16.2.11.3 Waveform Data
REFCurve<m>:DATA:STYPe..........................................................................................596
REFCurve<m>:DATA:HEADer........................................................................................596
REFCurve<m>:DATA[:VALues].......................................................................................597
REFCurve<m>:DATA:STYPe?
Returns the signal type of the source of the reference waveform.
Suffix:
<m>
.
1..4
Reference waveform
Return values:
<SignalType> SOUR | SPEC | CORR | NONE
SOURce = normal signal
SPECtrum = FFT spectrum, specific math signal
CORRelation = correlated signal, specific math signal
NONE = undefined
Usage: Query only
SCPI confirmed
REFCurve<m>:DATA:HEADer?
Returns information on the reference waveform.
Table 16-19: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 200000
4 Number of values per sample interval. For most
waveforms the result is 1, for peak detect and enve-
lope waveforms it is 2.
1
Suffix:
<m>
.
1..4
Reference waveform
Example:
REFC:DATA:HEAD?
-9.477E-008,9.477E-008,200000,1
Command Reference