User manual

Remote Control
R&S
®
RTO
593User Manual 1316.0827.02 ─ 06
Parameters:
<VerticalMode> COUPled | INDependent
COUPled
Vertical position and scale of the source are used.
INDependent
Scaling and position can be set specific to the reference waveform.
*RST: INDependent
REFCurve<m>:SCALe <VerticalScale>
Sets the scale factor for the reference waveform if vertical scaling is set to "Independ-
ent" (see REFCurve<m>:VMODe on page 592).
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<VerticalScale> Range: 100E-12 to 100E+24
Increment: 10E-6
*RST: 0.5
Default unit: V/div
Usage: SCPI confirmed
REFCurve<m>:POSition <VertPosi>
Moves the reference waveform and its horizontal axis up or down in the diagram, if vertical
scaling is set to "Independent" (see REFCurve<m>:VMODe on page 592).
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<VertPosi> Range: -100E+24 to 100E+24
Increment: 0.2
*RST: 0
Default unit: div
Usage: SCPI confirmed
REFCurve<m>:RESCale:VERTical:STATe <State>
Enables and disables the vertical stretching. Stretching changes the display of the wave-
form independent of the vertical scale and position.
Suffix:
<m>
.
1..4
Reference waveform
Command Reference