User manual
Remote Control
R&S
®
RTO
586User Manual 1316.0827.02 ─ 06
CALCulate:MATH<m>:FFT:GATE:ZCOupling <ZoomCoupling>
If enabled, the gate area is defined identically to the zoom area for the zoom diagram.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<ZoomCoupling> ON | OFF
*RST: OFF
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:GATE[:STATe] <State>
Enables FFT gating.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<State> ON | OFF
*RST: OFF
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:MAGNitude:LEVel <ReferenceLevel>
Defines the reference level for dB scaling.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<ReferenceLevel> Range: -100 to 50
Increment: 0.1
*RST: 0
Default unit: dBm
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:MAGNitude:RANGe <Range>
Defines the vertical value range in spectrum mode.
Suffix:
<m>
.
1...4
math waveform
Command Reference