User manual
Remote Control
R&S
®
RTO
585User Manual 1316.0827.02 ─ 06
Parameters:
<Mode> ABS | REL
*RST: ABS
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:GATE:RELative:STARt <RelativeStart>
Defines the starting value for the gate in percent.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<RelativeStart> Range: -100E+24 to 100E+24
Increment: 0.1
*RST: 0
Default unit: %
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:GATE:RELative:STOP <RelativeStop>
Defines the end value for the gate in percent.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<RelativeStop> Range: -100E+24 to 100E+24
Increment: 0.1
*RST: 100
Default unit: %
Usage: SCPI confirmed
CALCulate:MATH<m>:FFT:GATE:SHOW <DisplayState>
Indicates the gate area in the source diagram.
Suffix:
<m>
.
1...4
math waveform
Parameters:
<DisplayState> ON | OFF
*RST: OFF
Usage: SCPI confirmed
Command Reference