User manual
Remote Control
R&S
®
RTO
580User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..4
math waveform
Parameters:
<CenterFreq> center frequency
Range: 0 to 2E+12
Increment: 1
*RST: 2.5E+9
Default unit: Hz
CALCulate:MATH<m>:FFT:FULLspan
Performs FFT calculation for the full frequency span.
Suffix:
<m>
.
1..4
math waveform
Usage: Event
CALCulate:MATH<m>:FFT:SPAN <FreqSpan>
The span is specified in Hertz and defines the width of the displayed frequency range,
which is (Center - Span/2) to (Center + Span/2). The position of the span is defined using
the CALCulate:MATH<m>:FFT:CFRequency command.
Suffix:
<m>
.
1..4
Math waveform
Parameters:
<FreqSpan> Frequency span
Range: 100 to 4E+12
Increment: 100
*RST: 5E+9
Default unit: Hz
CALCulate:MATH<m>:FFT:BANDwidth[:RESolution]:ADJusted?
Queries the effective resolution bandwidth.
Suffix:
<m>
.
1..4
math waveform
Command Reference