User manual

Remote Control
R&S
®
RTO
576User Manual 1316.0827.02 ─ 06
CALCulate:MATH<m>:STATe <State>
Enables the math waveform display.
Suffix:
<m>
.
1..4
math waveform
Parameters:
<State> ON | OFF
*RST: OFF
Usage: SCPI confirmed
CALCulate:MATH<m>:ARIThmetics <Arithmetics>
Selects the method to build the resulting math waveform from consecutive acquisitions.
The processing is similar to the waveform arithmetics - instead of the acquired waveforms
the math waveforms are used to create envelope and average.
See also: CHANnel<m>[:WAVeform<n>]:ARIThmetics on page 434.
Suffix:
<m>
.
1..4
Selects the math waveform.
Parameters:
<Arithmetics> OFF | ENVelope | AVERage
waveform arithmetic mode
*RST: OFF
CALCulate:MATH<m>:VERTical:OFFSet <VerticalOffset>
Sets a voltage offset to adjust the vertical position of the math function on the screen.
Suffix:
<m>
.
1..4
math waveform
Parameters:
<VerticalOffset> Negative values move the waveform au, positive values move it
down.
Range: -100E+24 to 100E+24
Increment: 0.01
*RST: 0
Default unit: div
Usage: SCPI confirmed
CALCulate:MATH<m>:VERTical:RANGe <VerticalRange>
Defines the range of FFT values to be displayed.
Command Reference