User manual

Remote Control
R&S
®
RTO
569User Manual 1316.0827.02 ─ 06
MEASurement<m>:STATistics:WEIGht <Weight>
Defines the number of measurements for which the statistical average is calculated.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<Weight> Statistics average count
Range: 2 to 2147483647
Increment: 10
*RST: 10000
MEASurement<m>:
LTMeas[:STATe] <ShowDiagram>
Enables long term measurement for a defined number of measurement points (see
MEASurement<m>:LTMeas:COUNt on page 569) or a specified time (see
MEASurement<m>:LTMeas:TIME on page 569).
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<ShowDiagram> ON | OFF
*RST: OFF
MEASurement<m>:LTMeas:COUNt <MeasCount>
Defines the total number of points to be measured during the long term measurement.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<MeasCount> Range: 1000 to 200000
Increment: 10
*RST: 1000
MEASurement<m>:LTMeas:TIME <MeasurementTime>
Defines the total duration of the long term measurement.
This setting is only available if MEASurement<m>:STATistics:MODE is set to "Time".
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Command Reference