User manual

Remote Control
R&S
®
RTO
567User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<ResetMode> NONE | TIME | WFMS
NONE
No reset, the number of measurements considered by the statis-
tics is not limited.
Time
Resets the statistics after the time defined using
MEASurement<m>:STATistics:RTIMe.
Waveforms
Resets the statistics after a number of measurements defined
using MEASurement<m>:STATistics:RCOunt.
*RST: NONE
MEASurement<m>:
STATistics:RCOunt <RstWfmCount>
Defines the number of measurements after which the statistics are reset.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<RstWfmCount> Range: 1 to 65535
Increment: 10
*RST: 1000
MEASurement<m>:STATistics:RTIMe <ResetTime>
Defines the time or period after which the statistics are reset.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<ResetTime> Range: 0.1 to 2.14748E+9
Increment: 1E-3
*RST: 0.2
Default unit: s
MEASurement<m>:STATistics:RESet
Resets the statistics.