User manual

Remote Control
R&S
®
RTO
562User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<HistogramName> Name of the histogram
MEASurement<m>:HISTogram:PROBability:TYPE <Marker>
Defines the marker reference in the probability domain.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<Marker> PEAK | UPPK | LWPK | MAXimum | MINimum | MEDian | MEAN
PEAK
The y-value with the maximum sample value in the histogram
UPPK
The y-value at the maximum sample value in the upper half of the
histogram
LWPK
The y-value at the maximum sample value in the lower half of the
histogram
MAXimum
The highest y-value with a probability > 0
MINimum
The lowest y-value with a probability > 0
MEDian
The y-value for which half the samples lie above, the other half
below in the histogram
MEAN
The weighted arithmetic average of the histogram
*RST:
PEAK
MEASurement<m>:HISTogram:PROBability:LIMit <Limit>
Defines a range around the probability marker.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.