User manual
Remote Control
R&S
®
RTO
555User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..8
Number of spectrum measurement type, see table 16-16.
Parameters:
<LowerLimit>
<UpperLimit>
Range: -100 to 100
Increment: 10E-6
*RST: 0
MEASurement<m>:
SPECtrum:LCHeck<n>:LOWer:MARGin <LowerMargin>
MEASurement<m>:SPECtrum:LCHeck<n>:UPPer:MARGin <UpperMargin>
Define the lower and upper margins for the limit check, respectively. Margins are not as
strict as limits and must be within the valid value range. The valid range is defined using
the MEASurement<m>:SPECtrum:LCHeck<n>:VALid command.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..8
Number of spectrum measurement type, see table 16-16.
Parameters:
<Lower Margin>
<UpperMargin>
Range: -100 to 100
Increment: 10E-6
*RST: 0
16.2.9.6 Histograms
● Histogram Display.................................................................................................555
● Histogram Measurement.......................................................................................560
Histogram Display
LAYout:HISTogram:ADD................................................................................................556
LAYout:HISTogram:SOURce..........................................................................................556
LAYout:HISTogram:MODE..............................................................................................557
LAYout:HISTogram:HORZ:MODE...................................................................................557
LAYout:HISTogram:HORZ:ABSolute:STARt.....................................................................557
LAYout:HISTogram:HORZ:ABSolute:STOP......................................................................558
LAYout:HISTogram:HORZ:RELative:STARt......................................................................558
LAYout:HISTogram:HORZ:RELative:STOP......................................................................558
LAYout:HISTogram:VERTical:MODE...............................................................................558
LAYout:HISTogram:VERTical:ABSolute:STARt.................................................................559
LAYout:HISTogram:VERTical:ABSolute:STOP..................................................................559
LAYout:HISTogram:VERTical:RELative:STARt.................................................................559
LAYout:HISTogram:VERTical:RELative:STOP..................................................................560
LAYout:HISTogram:RESet..............................................................................................560
Command Reference