User manual
Remote Control
R&S
®
RTO
554User Manual 1316.0827.02 ─ 06
Parameters:
<NDbDown> Range: 0 to 100
Increment: 1
*RST: 20
Default unit: dB
MEASurement<m>:SPECtrum:NREJect <NoiseReject>
Defines the threshold beneath which values are rejected as noise.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<NoiseReject> Range: 0 to 100
Increment: 1
*RST: 0
Default unit: dB
MEASurement<m>:SPECtrum:LCHeck<n>:VALid <ValidRange>
Enables or disables limit checking for spectrum measurements in the specified mea-
surement channel.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..8
Number of spectrum measurement type, see table 16-16.
Parameters:
<ValidRange> ILIMit | ULIMit | LLIMit | OLIMit
ILIMit
Inside (within) limit; between the upper and lower limit values
ULIMit
Upper limit; above the upper limit value
LLIMit
Lower limit; below the lower limit value
OLIMit
Outside limit; above the upper limit or below the lower limit values
*RST: ILIMit
MEASurement<m>:SPECtrum:LCHeck<n>:LOWer:LIMit <LowerLimit>
MEASurement<m>:SPECtrum:LCHeck<n>:UPPer:LIMit <UpperLimit>
Define the lower and upper limits for the limit check, respectively. The valid range is
defined using the MEASurement<m>:SPECtrum:LCHeck<n>:VALid command.
Command Reference