User manual

Remote Control
R&S
®
RTO
551User Manual 1316.0827.02 ─ 06
MEASurement<m>:EYEJitter:LCHeck<n>:VALid <ValidRange>
Enables or disables limit checking for eye/jitter measurements in the specified measure-
ment channel.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..30
Number of eye/jitter measurement type, see table 16-15.
Parameters:
<ValidRange> ILIMit | ULIMit | LLIMit | OLIMit
ILIMit
Inside (within) limit; between the upper and lower limit values
ULIMit
Upper limit; above the upper limit value
LLIMit
Lower limit; below the lower limit value
OLIMit
Outside limit; above the upper limit or below the lower limit values
*RST: ILIMit
MEASurement<m>:EYEJitter:LCHeck<n>:LOWer:LIMit <LowerLimit>
MEASurement<m>:EYEJitter:LCHeck<n>:UPPer:LIMit <UpperLimit>
Define the lower and upper limit for the limit check, respectively. The valid range is defined
using the MEASurement<m>:EYEJitter:LCHeck<n>:VALid command.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..30
Number of eye/jitter measurement type, see table 16-15.
Parameters:
<LowerLimit>
<UpperLimit>
Range: -100 to 100
Increment: 10E-6
*RST: 0
MEASurement<m>:EYEJitter:LCHeck<n>:LOWer:MARGin <LowerMargin>
MEASurement<m>:EYEJitter:LCHeck<n>:UPPer:MARGin <UpperMargin>
Defines the upper margin for the limit check. Margins are not as strict as limits and must
be within the valid value range. The valid range is defined using the
MEASurement<m>:EYEJitter:LCHeck<n>:VALid command.
Command Reference