User manual

Remote Control
R&S
®
RTO
548User Manual 1316.0827.02 ─ 06
Parameters:
<SetupHoldClockSlope>POSitive | NEGative | EITHer
EITHer
The clock edges next to the data edge are considered regardless
of the clock slope.
*RST: POSitive
MEASurement<m>:AMPTime:CLCK<n>:LSELect <ClockLevel>
Selects the reference level of the clock on which the time is measured. Reference level
and clock slope define the time point for setup and hold measurements.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..2
The suffix is irrelevant, omit it.
Parameters:
<ClockLevel> UPPer | MIDDle | LOWer
*RST: MIDDle
MEASurement<m>:AMPTime:DATA<n>:LSELect <DataLevel>
Selects the reference level of the data on which the setup and hold time are measured.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..2
The suffix is irrelevant, omit it.
Parameters:
<DataLevel> UPPer | MIDDle | LOWer
*RST: MIDDle
MEASurement<m>:AMPTime:LCHeck<n>:VALid <ValidRange>
Enables or disables limit checking for amplitude vs. time measurements in the specified
measurement channel.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
<n> 1..37
Number of the amplitude/time measurement type, see
table 16-14.
Command Reference