MEASurement<m>:DETThreshold <SignDetectThres>
Defines the value above which measurement results are displayed. Values beneath the
threshold are considered to be noise and they are ignored.
Suffix:
<m>
.
1..9
irrelevant
Parameters:
<SignDetectThres> Range: 0 to 50
Increment: 1
*RST: 5
Default unit: %
MEASurement<m>:
AMPTime:ALEVel <AreaLevel>
Defines the reference level used to integrate the waveform.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<AreaLevel> Range: -100E+24 to 100E+24
Increment: 0
*RST: 0
Default unit: V
MEASurement<m>:AMPTime:PSLope <PulsesSlope>
Sets the first slope of the pulses to be counted. The setting is only relevant for pulse count
measurement (MEASurement<m>:MAIN PULCnt or
MEASurement<m>:ADDitional PULCnt,ON).
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<PulsesSlope> POSitive | NEGative | EITHer
Count either positive or negative pulses, or both.
*RST: POSitive
MEASurement<m>:AMPTime:DELay<n>:DIRection <EdgeCountDirection>
Selects the direction for counting slopes for each source: from the beginning of the wave-
form, or from the end.