User manual

Remote Control
R&S
®
RTO
542User Manual 1316.0827.02 ─ 06
Return values:
<TopOuter> Range: -100E+24 to 100E+24
Increment: 0
*RST: 0
Usage: Query only
SCPI confirmed
16.2.9.3 Amplitude/Time Measurement
The following table lists the measurement suffixes and the <MeasType> parameter value
with a short description.
For a detailed description, see chapter 5.2.1.2, "Amplitude/Time Measure-
ments", on page 132.
Table 16-14: Amplitude and time measurement types
Meas.
suffix
<MeasType>
parameter value
Meas. type Description/Result
1 HIGH High High signal level
2 LOW Low Low signal level
3 AMPLitude Amplitude Amplitude of the signal
4 MAXimum Max Maximum value of the waveform
5 MINimum Min Minimum value of the waveform
6 PDELta Peak to peak Peak-to-peak value of the waveform
7 MEAN Mean Mean value of the waveform
8 RMS RMS RMS (Root Mean Square) value of the voltage
9 STDDev
σ (S-dev)
Standard deviation of the waveform
10 POVershoot Pos. overshoot Positive overshoot of a square wave
11 NOVershoot Neg. overshoot Negative overshoot of a square wave
12 AREA Area Area beneath the waveform (integral)
13 RTIMe Rise time Rise time of the left-most rising edge of the waveform.
14 FTIMe Fall time Falling time of the left-most falling edge of the waveform.
15 PPULse Pos. pulse
Width of a positive pulse a rising edge followed by a falling
edge. The measurement requires at least one complete
period of a triggered signal.
16 NPULse Neg. pulse
Width of a negative pulse a falling edge followed by a
rising edge. The measurement requires at least one com-
plete period of a triggered signal.
17 PERiod Period Length of the left-most signal period of the waveform
18 FREQuency Frequency Frequency of the signal. The result is based on the period
measurement.