User manual
Remote Control
R&S
®
RTO
530User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
2...21
Source waveform, see chapter 16.2.2.1, "Waveform Suf-
fix", on page 419.
Parameters:
<UserLevel> USIGnal | UREF
USIGnal
The high and low signal levels are defined by the user.
UREF
The reference levels are defined by the user.
*RST: USIGnal
Example:
REFLevel2:USRLevel UREF
Sets user-defined reference levels to be used for Channel1/Wave-
form1. C1W1 corresponds to suffix number 2.
See also: example "Manual reference level definition using abso-
lute values" on page 527
REFLevel<m>:LMODe <LevelMode>
Defines whether the reference is configured using absolute or relative values.
Suffix:
<m>
.
2...21
Source waveform, see chapter 16.2.2.1, "Waveform Suf-
fix", on page 419.
Parameters:
<LevelMode> ABS | REL
*RST: REL
Example:
REFLevel2:LMODe ABS
Sets definition of reference levels to absolute values for Channel1/
Waveform1. C1W1 corresponds to suffix number 2.
See also: example "Manual reference level definition using relative
values" on page 528
Automatic Configuration
REFLevel<m>:AUTO:MODE...........................................................................................530
REFLevel<m>:AUTO[:STATe].........................................................................................531
REFLevel<m>:AUTO:COUNt..........................................................................................532
REFLevel<m>:AUTO:MODE <AutoLevelMode>
Defines the high and low signal levels from which the reference levels are derived.
This setting is only available for automatic reference level mode (see REFLevel<m>:
LDETection on page 528).
Command Reference