User manual

Remote Control
R&S
®
RTO
527User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Usage: Event
MEASurement<m>:MULTiple <MultiMeas>
The measurement is performed repeatedly if the measured parameter occurs several
times inside the defined gate.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<MultiMeas> ON | OFF
*RST: OFF
MEASurement<m>:MNOMeas <MaxNoOfMeasPerAcq>
Sets the maximum number of measurements per acquisition if multiple measurement is
enabled (MEASurement<m>:MULTiple is ON.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<MaxNoOfMeasPerAcq>Range: 2 to 100000
Increment: 1
*RST: 1000
16.2.9.2 Reference Level
Programming Examples........................................................................................527
General Reference Level Settings........................................................................528
Automatic Configuration........................................................................................530
Manual Configuration............................................................................................532
Hysteresis.............................................................................................................537
Tube......................................................................................................................537
Results..................................................................................................................540
Programming Examples
Example: Manual reference level definition using absolute values
The modes, the upper and lower reference level, and the top and bottom distance are
set for waveform C1W1.
Command Reference