User manual
Remote Control
R&S
®
RTO
525User Manual 1316.0827.02 ─ 06
Parameters:
<MeasType> Amplitude/time measurements
HIGH | LOW | AMPLitude | MAXimum | MINimum | PDELta | MEAN
| RMS | STDDev | POVershoot | NOVershoot | AREA | RTIMe |
FTIMe | PPULse | NPULse | PERiod | FREQuency | PDCYcle |
NDCYcle | CYCarea | CYCMean | CYCRms | CYCStddev |
PULCnt | DELay | PHASe | BWIDth | PSWitching | NSWitching |
PULSetrain | EDGecount | SHT | SHR | PROBemeter
See chapter 16.2.9.3, "Amplitude/Time Measure-
ment", on page 542.
*RST value for amplitude/time measurements: AMPLitude.
Eye/jitter measurements
ERPercent | ERDB | EHEight | EWIDth | ETOP | EBASe | QFACtor
| RMSNoise | SNRatio | DCDistortion | ERTime | EFTime | EBRate
| EAMPlitude | PPJitter | STDJitter | RMSJitter
See chapter 16.2.9.4, "Eye/Jitter Measurement", on page 550.
*RST value for eye/jitter measurements: ERPercent.
Spectrum measurements
CPOWer | OBWidth | SBWidth | THD
See chapter 16.2.9.5, "Spectrum", on page 552.
*RST value spectrum measurements: CPOWer.
Histogram measurements
WCOunt | WSAMples | HSAMples | HPEak | PEAK | UPEakvalue
| LPEakvalue | HMAXimum | HMINimum | MEDian | MAXMin |
HMEan | HSTDdev | M1STddev | M2STddev | M3STddev |
MKPositive | MKNegative
See "Histogram Measurement" on page 560.
*RST value for histogram measurements: WCOunt.
MEASurement<m>:ADDitional <MeasType>, <State>
MEASurement<m>:ADDitional? <MeasType>
Enables or disables an additional measurement. Only one measurement type can be
enabled or disabled per command. The query returns the state of the specified mea-
surement type.
Note that each measurement waveform can only perform measurements from the same
category. For example, if you enable an amplitude measurement for measurement wave-
form 1, then you cannot enable an eye width measurement for the same waveform.
For details on the measurement types and categories, see chapter 5.2.1, "Measurement
Types and Results", on page 129.
Suffix:
<m>
.
1..9
See "Measurement selection: MEASurement<m>" on page 522.
Parameters:
<State> ON | OFF
Enables or disables the measurement type.
Command Reference