Enable ultra segmentation
Switches the Ultra Segmentation mode on and off.
If "Equivalent time" sampling is selected in the "Acquisition" tab, enabling Ultra Segmen-
tation switches the resolution enhancement to "Interpolated time".
SCPI command:
ACQuire:SEGMented:STATe on page 435
Number of acquisitions
You can define the number of acquisitions to stored in an Ultra Segmentation acquisition
series:
Acquire the maximum possible number of acquisitions that can be stored in the sam-
ple memory.
To acquire the maximum number, enable "Acquire maximum". The maximum number
of acquisitions is shown in the "Required" field.
Acquire a given number of acquisitions.
Enter the number in the "Required" field.
The acquisition count (Acquisition/average count) is always set to the required number
of acquisitions. Thus you can acquire exactly one Ultra Segmentation acquisition series
with RUN N×SINGLE. The RUN key works in the same way as RUN N×SINGLE, it stops
acquisition when the series is completed.
You can stop the running acquisition before the series is completed.
The number of actually acquired waveforms is shown in "Available" and can be displayed
with "Show history".
SCPI command:
ACQuire:SEGMented:MAX on page 435
Replay time
Defines the display speed of the Ultra Segmentation acquisition series. Display starts
after the series has been captured completely. See also "Time per acquisi-
tion" on page 119.
Reference for Acquisition and Setup