User manual

Remote Control
R&S
®
RTO
513User Manual 1316.0827.02 ─ 06
Parameters:
<StopAcqIdx> Number of the stop acquisition. The newest acquisition always has
the index "0".
Range: 0 to 0
Increment: 1
*RST: 0
CHANnel<m>[:WAVeform<n>]:HISTory:TPACq <TimePerAcq>
Sets the display time for one acquisition. The shorter the time, the faster is the replay.
Suffix:
<m>
.
1..4
Selects the input channel.
<n> 1..3
Selects the waveform. For each channel, up to three waveforms
can be analyzed. If WAVeform<n> is omitted, waveform 1 is
adressed.
Parameters:
<TimePerAcq> Range: 40E-6 to 10
Increment: 1
*RST: 0.05
Default unit: s
CHANnel<m>[:WAVeform<n>]:HISTory:PLAY
Starts and stops the replay of the history waveforms.
Suffix:
<m>
.
1..4
Selects the input channel.
<n> 1..3
Selects the waveform. For each channel, up to three waveforms
can be analyzed. If WAVeform<n> is omitted, waveform 1 is
adressed.
Usage: Event
Asynchronous command
CHANnel<m>[:WAVeform<n>]:HISTory:REPLay <AutoRepeat>
If ON, the replay of the history waveform sequence repeats automatically. Otherwise, the
replay stops at the stop index set with CHANnel<m>[:WAVeform<n>]:HISTory:
STOP.
Suffix:
<m>
.
1..4
Selects the input channel.
Command Reference