User manual
Remote Control
R&S
®
RTO
492User Manual 1316.0827.02 ─ 06
Return values:
<Delay>
Constant value: 250 ns
Range: 1E-9 to 1
Increment: 1E-9
*RST: 800E-9
Default unit: s
Usage: Query only
16.2.7
Display
● Signal Colors / Persistence...................................................................................492
● Color Tables..........................................................................................................494
● Diagram Layout.....................................................................................................496
● Zoom.....................................................................................................................502
● XY-Diagram...........................................................................................................509
● History...................................................................................................................511
16.2.7.1 Signal Colors / Persistence
DISPlay:PERSistence[:STATe]........................................................................................492
DISPlay:PERSistence:INFinite.........................................................................................492
DISPlay:PERSistence:TIME............................................................................................493
DISPlay:PERSistence:RESet..........................................................................................493
DISPlay:INTensity..........................................................................................................493
DISPlay:DIAGram:STYLe...............................................................................................493
DISPlay:COLor:SIGNal<m>:ASSign.................................................................................494
DISPlay:COLor:SIGNal<m>:USE.....................................................................................494
DISPlay:PERSistence[:STATe] <State>
If enabled, each new data point in the diagram area remains on the screen for the duration
defined using DISPlay:PERSistence:TIME, or as long as DISPlay:
PERSistence:INFinite is enabled.
If disabled, the signal value is only displayed as long as it actually occurs.
Parameters:
<State> ON | OFF
*RST: ON
DISPlay:PERSistence:INFinite <State>
If persistence is enabled (DISPlay:PERSistence[:STATe]), each new data point in
the diagram area remains on the screen infinitely until this command is set to "OFF".
Command Reference