User manual

Remote Control
R&S
®
RTO
489User Manual 1316.0827.02 ─ 06
Parameters:
<AutoTimeScaling> Range: 1E-3 to 1000
Increment: 1
*RST: 0.5
16.2.6.5 Trigger Position
TIMebase:REFerence.....................................................................................................489
TIMebase:POSition........................................................................................................489
TRIGger<m>:OFFSet:LIMited..........................................................................................489
TIMebase:REFerence <ReferencePoint>
Sets the reference point of the time scale in % of the display. The reference point defines
which part of the waveform is shown. If the "Trigger offset" is zero, the trigger point
matches the reference point.
See also: TIMebase:POSition on page 429
Parameters:
<ReferencePoint> The reference point is the zero point of the time scale.
Range: 0 to 100
Increment: 1
*RST: 50
Default unit: %
TIMebase:POSition <Offset>
Defines the trigger offset - the time interval between trigger point and reference point to
analize the signal some time before or after the trigger event.
See also: TIMebase:REFerence on page 430
Parameters:
<Offset> Range: -500 to 500
Increment: 0.01
*RST: 0
Default unit: s
TRIGger<m>:OFFSet:LIMited <State>
Ensures that the trigger occurs within within one acquisition cycle.
Suffix:
<m>
.
1..3
The numeric suffix is irrelevant.
Command Reference