488User Manual 1316.0827.02 ─ 06
Suffix:
<m>
.
1..3
For holdoff settings, only suffix 1 (A-event) is available.
Parameters:
<RandomMaxTime> Range: 100E-9 to 10
Increment: 200E-6
*RST: 2E-3
Default unit: s
TRIGger<m>:HOLDoff:AUTotime?
Returns the resulting holdoff time if the holdoff mode is set to AUTO: Auto time = Auto
time scaling * Horizontal scale. The auto time scaling factor is defined with
TRIGger<m>:HOLDoff:SCALing.
See also: TRIGger<m>:HOLDoff:MODE
Suffix:
<m>
.
1..3
For holdoff settings, only suffix 1 (A-event) is available.
Return values:
<AutoTime> Holdoff time
Range: 100E-9 to 10
Increment: 200E-6
*RST: 1E-3
Default unit: s
Example:
TRIGger1:HOLDoff:MODE AUTO
TRIGger1:HOLDoff:SCALing 0.5
TRIGger<m>:HOLDoff:AUTotime?
1ms
Result if the horizontal scale is 1 ns/div
Usage: Query only
TRIGger<m>:HOLDoff:SCALing <AutoTimeScaling>
Sets the auto time scaling factor the horizontal scale is multipied with: Auto time = Auto
time scaling * Horizontal scale. The setting is relevant if the holdoff mode is set to AUTO.
See also:
● TRIGger<m>:HOLDoff:MODE
● TRIGger<m>:HOLDoff:AUTotime on page 488
Suffix:
<m>
.
1..3
For holdoff settings, only suffix 1 (A-event) is available.
Command Reference