Suffix:
<m>
.
1..3
For holdoff settings, only suffix 1 (A-event) is available.
Parameters:
<Mode> TIME | EVENts | RANDom | AUTO | OFF
TIME
Defines the holdoff directly as a time period. The next trigger
occurs only after the "Holdoff time" has passed (defined using
TRIGger<m>:HOLDoff:TIME).
EVENts
Defines the holdoff as a number of trigger events. The next trigger
occurs only when this number of events is reached. The number
of triggers to be skipped is defined using TRIGger<m>:
HOLDoff:EVENts.
RANDom
Defines the holdoff as a random time limited by TRIGger<m>:
HOLDoff:MIN on page 487 and TRIGger<m>:HOLDoff:MAX
on page 487. For each acquisition cycle, the instrument selects a
new random holdoff time from the specified range.
AUTO
The holdoff time is calculated automatically based on the current
horizontal scale.
OFF
No holdoff
*RST:
OFF
TRIGger<m>:HOLDoff:TIME <Time>
Defines the holdoff time period. The next trigger occurs only after this time has passed.
The setting is relevant if the holdoff mode is set to TIME.
See also:
TRIGger<m>:HOLDoff:MODE
Suffix:
<m>
.
1..3
For holdoff settings, only suffix 1 (A-event) is available.
Parameters:
<Time> Range: 100E-9 to 10
Increment: 200E-6
*RST: 1E-3
Default unit: s
Example:
TRIGger1:HOLDoff:MODE TIME
TRIGger<m>:HOLDoff:TIME 1ms
The holdoff time is set to 1 ms.