485User Manual 1316.0827.02 ─ 06
TRIGger<m>:SEQuence:RESet:EVENt <EnabRstEvt>
If set to ON, the trigger sequence is restarted by the R-event if the specified number of
B-event does not occur.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<EnabRstEvt> ON | OFF
*RST: OFF
TRIGger<m>:SEQuence:RESet:TIMeout[:ENABle] <State>
If set to ON, the instrument waits for the time defined using TRIGger<m>:SEQuence:
RESet:TIMeout:TIME for the specified number of B-events. If no trigger occurs during
that time, the sequence is restarted with the A-event.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<State> ON | OFF
*RST: OFF
TRIGger<m>:SEQuence:RESet:TIMeout:TIME <ResetTimeout>
The time the instrument waits for the number of B-events specified using
TRIGger<m>:SEQuence:COUNt before the sequence is restarted with the A-event.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<ResetTimeout> Range: 0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
TRIGger<m>:HOLDoff:MODE <Mode>
Selects the method to define the holdoff condition.
The trigger holdoff defines when the next trigger after the current will be recognized. Thus,
it affects the next trigger to occur after the current one. Holdoff helps to obtain stable
triggering when the oscilloscope is triggering on undesired events.
Command Reference